Product advantages Plinma Disposable Non-invasive EEG Sensor is easy to use, safe and non-invasive, and ofered in multi specications for both adults and children, including 86A03, 86A04, 86P03, 86P04 and etc....
Product advantages
Plinma Disposable Non-invasive EEG Sensor is easy to use, safe and non-invasive, and ofered in multi specications for both adults and children, including 86A03, 86A04, 86P03, 86P04 and etc.
Through foreign clinical verification, the sensor signal is stable, and the response is fast. Compared with the original product, the numerical consistency is good. It can accurately reflect the patient's current state of depth of anesthesia.
Double shielding technology is used to reduce impedance and reduce interference. Imported conductive glue and sensor components are used to ensure complete and accurate data collection and real-time monitoring.
Various materials of the product have passed rigorous biocompatibility tests. No sensitization.
Can be adapted to multi-brand anesthesia depth monitoring equipment (BIS module and GE entropy index module)
Technical parameter
AC impedance | ≤3KΩ(10Hz) |
DC offset voltage | ≤100mV |
Internal noise | ≤150μV(peak-peak) |
Simulated defibrillation recovery performance | The voltage value of the electrode pair at 5s after discharge is less than or equal to 100mV; the average rate of change after 30s is less than or equal to ±1mV/s |
Simulated AC impedance after defibrillation | ≤3KΩ |
Bias current tolerance | ≤100mV |
Cytotoxicity test | The cytotoxicity score is not greater than 1 |
Delayed type hypersensitivity test | No sensitization |